Place of Origin: | CHINA |
---|---|
Brand Name: | AI |
Certification: | ISO |
Model Number: | AIC-6049 |
Minimum Order Quantity: | 1 Set Low Temperature Impact Test Machine |
Price: | Negotiable |
Packaging Details: | Low Temperature Impact Test Machine packed in Hardboard |
Delivery Time: | 20 days |
Payment Terms: | L/C, T/T, Western Union |
Supply Ability: | 100 sets per month |
Test Standard: | IEC 60811-1-4 Clause 8 | Impact Height: | 100mm |
---|---|---|---|
Impact Diameter: | 20mm | Radian Of Impact Head: | 300mm |
Dimension: | 37×15×44cm | Weight: | 23KG |
Weights: | 100g,200g,300g,400g,500g,600g,750g,1000g,1250g,1500g | Height Of Base: | 40mm |
Highlight: | Low Temperature Impact Test Machine,Impact Test Machine for Cable,IEC 60811-1-4 |
Application
When using this impact test equipment,it should be placed in low temperature chamber with specimen.After cooling,the weight with specified weight drops from 100mm high above and impact the specimen,then observe the damage of inner and outer margins of isolation and sheath.
This equipment should be used with AIC-6034 low temperature test chamber.
Reference standards:
● GB2951.14-2008
● IEC60811-1-4:1985
Main Specifications:
1 | Impact height | 100mm |
2 | Impact diameter | 20mm |
3 | Radian of impact head | 300mm |
4 | Guiding rod of impact hammer | Ф10mm,length 8.6cm,equilateral triangle |
5 | Weights | 100g,200g,300g,400g,500g,600g,750g,1000g,1250g,1500g,totally 10 pcs/group |
6 | Base weight of impact device | 10KG |
7 | Height of base | 40mm |
8 | Dimension | 37×15×44cm |
9 | Weight | 23KG |